Table 3 Comparison of analyte content in samples, determined by two methods

From: Development of an HPLC method using relative molar sensitivity for the measurement of blood concentrations of nine pharmaceutical compounds

 

Sample name

Spiked

concentration

(μg/mL)

RMS method

(Calibrant: Carbamazepine)

RMS method

(Calibrant: Caffeine)

Conventional method

Concentration (μg/mL)

RSD

(%)

Concentration (μg/mL)

RSD

(%)

Concentration (μg/mL)

RSD

(%)

Carbamazepine spiked

Sample 1

20

21.1

1.9

21.1

1.9

21.1

1.9

Phenytoin spiked

Sample 2

2

2.1

0.9

2.1

0.9

2.1

0.9

Sample 3

20

18.0

2.5

18.1

2.5

18.1

2.5

Voriconazole spiked

Sample 4

1

1.1

1.0

1.0

1.1

1.1

1.0

Sample 5

5

5.1

2.0

5.2

1.9

5.2

1.9

Meropenem spiked

Sample 6

0.5

0.4

1.4

0.4

1.4

0.4

1.4

Sample 7

100

85.5

2.8

85.9

2.8

85.5

2.8

Mycophenolic acid spiked

Sample 8

1

1.1

0.5

1.1

0.5

1.1

0.5

Sample 9

20

17.5

1.3

17.5

1.3

17.5

1.3

Linezolid spiked

Sample 10

2

2.3

0.7

2.3

0.7

2.3

0.7

Sample 11

50

54.9

5.5

54.8

5.5

54.7

5.5